Fujifilm FDR D-EVO™ GL Brochure

Page de 2
1
FDR D-EVO GL
CR Cassette
Conventional method
ISS system reading technology
X-ray
X-ray
Scintillation
layer
TFT detector
Optical signal image
reaching TFT detector
Optical signal image
reaching TFT detector
TFT detector
Scintillation
layer
Blending optimally-sized phosphor particles in gaps
1
2
3
cm
43.2 
 
124.5
Long length Flat Panel Detector
Conventional DR system
FDR D-EVO GL
FDR D-EVO GL allows long-view radiography of
the entire lower limb or vertebral column with a single exposure.
43.2 
 124.5 cm size Flat Panel Detector
FDR D-EVO GL is approximately the size of three conventional 
43.2 
 43.2 cm [17 
 17 inch] DR detectors, with a wide field of 
view to accommodate all patient sizes without compromising 
the length of the acquisition.
Long-view radiography on DR systems 
traditionally requires multiple exposures, 
dramatically increasing the time the patient 
must hold still compared to CR systems. 
Using one exposure with the FDR D-EVO GL 
reduces time for the patient to remain 
steady; making the examination less 
susceptible to patient movement artifacts.
ISS capture technology promotes 
high sensitivity 
Fujifilm’s exclusive technologies for 
achieving low dose and sharp image
Improving sensitivity of 
low-density are as with 
a noise reduction circuitry
Image processing technology to 
optimize imaging results
Equipped with Fujifilm’s proprietary Irradiated Side Sampling (ISS) 
technology, which positions its capture electronics (TFTs) at the 
irradiation side, in contrast to traditional detectors. This design 
significantly suppresses scattering and attenuation of X-ray signals, 
improving efficiency to produce sharper images at lower doses 
compared to Conventional method.
The system uses image processing technologies to express the 
imaging results optimally, including Virtual Grid, which enhances 
image contrast and improves granularity degraded by scattered 
radiation, Dynamic Visualization, which provides images that are 
optimal for diagnoses on the monitor, and FNC noise suppression, 
which improves granularity by automatically extracting and 
separating noise components in the image.
The uniquely developed noise 
reduction circuitry reduces noise in the image.
1.
2.
3.
4.
Increased X-ray absorption through 
our uniquely designed scintillator, 
which applies photographic film 
technology to optimize the blending 
ratio of phosphor particles of 
different sizes.
Blending large and small phosphor particles at 
an optimal ratio
Fujifilm developed a new technology “SmartSwitch” which allows 
automatic X-ray detection. With “SmartSwitch,” FDR D-EVO GL no 
longer requires connection between the X-ray generator and DR power 
supply unit to automatically detect X-rays and start image creation.
“SmartSwitch” Technology
FDR D-EVO GL improves the efficiency of long-view exams, 
reducing exam times for the patient.
World’s First
*
A single exposure is performed within a timeframe of milliseconds; 
significantly reducing the risk of patient movement compared to a 
traditional DR multi-shot acquisition over a period of several seconds.
*According to research conducted by 
FUJIFILM Corporation in March 2016
[17 
 49 inch] 
[17 
 49 inch] 
35.4
 
 
124.5 cm field of view
 [14 
 49 inch] 
<現在>
17x17 inch
<変更希望>
43.2x43.2 cm [17x17 inch]