Important safeguards/notices for use 
in the medical environments
1. All the equipments connected to this unit 
shall be certified according to Standard 
IEC60601-1, IEC60950, IEC60065 or 
other IEC/ISO Standards applicable to 
the equipments.
2. When this unit is used together with other 
equipment in the patient area*, the 
equipment shall be either powered by an 
isolation transformer or connected via an 
additional protective earth terminal to 
system ground unless it is certified 
according to Standard IEC60601-1.
3. The leakage current could increase when 
connected to other equipment.
4. This equipment generates, uses, and can 
radiate frequency energy. If it is not 
installed and used in accordance with the 
instruction manual, it may cause 
interference to other equipment. If this 
unit causes interference (which can be 
determined by unplugging the power 
cord from the unit), try these measures: 
Relocate the unit with respect to the 
susceptible equipment. Plug this unit and 
the susceptible equipment into different 
branch circuit. Consult your dealer.
  (According to Standard EN60601-1-2 
and CISPR11, Class B, Group 1)
IMPORTANT
Nameplate is located on the bottom.
R 1.5 m
* Patient Area
Caution
When you dispose of the unit or 
accessories, you must obey the law in the 
relative area or country and the regulation 
in the relative hospital.