Caution
When you dispose of the unit or accessories, you must 
obey the law in the relative area or country and the 
regulation in the relative hospital.
Warning on power connection
Use a proper power cord for your local power supply.
1. Use the approved Power Cord (3-core mains lead) / 
Appliance Connector / Plug with earthing-contacts that 
conforms to the safety regulations of each country if 
applicable.
2. Use the Power Cord (3-core mains lead) / Appliance 
Connector / Plug conforming to the proper ratings 
(Voltage, Ampere).
If you have questions on the use of the above Power Cord 
/ Appliance Connector / Plug, please consult a qualified 
service personnel.
Warning on power connection for medical 
use
Please use the following power supply cord.
With connectors (plug or female) and cord types other than 
those indicated in this table, use the power supply cord that 
is approved for use in your area.
*Note: Grounding reliability can only be achieved when the equipment is 
connected to an equivalent receptacle marked ‘Hospital Only’ or ‘Hospi-
tal Grade’.
For the customers in Europe
The manufacturer of this product is Sony Corporation, 1-
7-1 Konan, Minato-ku, Tokyo, Japan.
The Authorized Representative for EMC, medical devices, 
and product safety is Sony Deutschland GmbH, 
Hedelfinger Strasse 61, 70327 Stuttgart, Germany.
For any service or guarantee matters, please refer to the 
addresses given in separate service or guarantee 
documents.
Recommended separation distances between portable and mobile RF communications equipment and the 
UP-DR80MD
The UP-DR80MD is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. 
The customer or the user of the UP-DR80MD can help prevent electromagnetic interference by maintaining a minimum 
distance between portable and mobile RF communications equipment (Transmitters) and the UP-DR80MD as 
recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of transmitter 
W
Separation distance according to frequency of transmitter 
m
150 kHz to 80 MHz
d = 1.2
 √P
80 MHz to 800 MHz
d = 1.2 
P
800 MHz to 2.5 GHz
d = 2.3 
P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated a maximum output power not listed above, the recommended separation distance d in meters (m) can 
be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating 
of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and 
reflection from structures, objects and people.
United States and Canada
Plug Type
HOSPITAL GRADE*
Cord Type
Min.Type SJT
Min.18 AWG
Minimum Rating for Plug 
and Appliance Couplers
10A/125V
Safety Approval
UL Listed and CSA