4-2     Calibration and Adjustments
Rosemount Analytical Inc.    A Division of Emerson Process Management
Model 7003M
5.  Press ENTER to return to the RUN
mode.
6.  If the ratio between the
concentration value entered and
the sensor voltage is excessive, an
error message will be flashed.
7.  If the sample gas and the
calibration gas are each at constant
pressure, continue with Section 4-
2c on page 4-4 and enter the
pressure of each.
Oxygen
20.946%
±0.002
Nitrogen
78.084%
±0.004
Argon
0.934%
±0.004
CO
2
0.033%
±0.001
Water Vapor
0.000%
Other Gases
Balance
760 mm Hg; 0% Relative Humidity
1
Table 4-2.
Composition of Dry Atmospheric Air
PERCENT COMPOSITION ( .002%) A@
GAS
15
°C
20
°C
25
°C
30
°C
Oxygen
20.641%
20.510%
20.338%
20.115%
Water Vapor
1.683
2.307
3.126
4.187
Other Gases
Balance
Balance
Balance
Balance
760 mm Hg; 100% Relative Humidity
2
Table 4-3.
Composition of Humid Atmospheric Air
                                                     
1
 CRC Handbook of Chemistry and Physics, R. Weast, ed., Chemical Rubber Co., Boca Raton, FL:  CRC Press, 1987, p. F-
148.
2
 Computed from data in CRC Handbook of Chemistry and Physics, p. D-190.