Calibration and Adjustments      4-5
Model 7003M
4-3 HOLD MODE
Refer to Figure 4-2 on page 4-7 and Table 4-6
below.  In the HOLD mode:
• 
All instrument functions are locked
• 
All existing values are retained
• 
The instrument stops responding to
sensor signals
• 
The last measured value is displayed
• 
The last output value remains at the
isolated current output connections
• 
Alarms and ON/OFF-Controllers are held
in their last state
• 
No keys are recognized except entry to
the diagnostic routing.
KEY
DISPLAY
WHAT HAPPENS
DISPLAY
Nothing
Nothing
ACK
Nothing
Nothing
ALARM 1
Nothing
Nothing
ALARM 2
Nothing
Nothing
UP ARROW
Nothing
Nothing
DOWN ARROW
Nothing
Nothing
CAL
Nothing
Nothing
ENTER
"run"
Drops back to RUN mode
DISPLAY + ACK
"d-4" "tSt"
Moves to DIAGNOSTIC mode
Table 4-6.
HOLD  Mode Key Functions
If the sensor cable must be disconnected, the
instrument can be put in the HOLD mode, in
which all existing signal levels and output
current levels and output current levels are
maintained.  This may be useful to prevent
disturbing an ON/OFF-control loop connected
to the relay contacts.
1.  Press the UP ARROW and DOWN
ARROW at the same time to put the
instrument on HOLD.
2.  If the security is activated, enter the
security identification (id) code and press
ENTER to continue.
3.  Hld, YES, flash briefly, then Hld flashes
alternately with the last oxygen reading.
The output signal will remain unchanged
until the instrument returns to the RUN
mode.
4.  Press ENTER to return to the RUN mode,
or continue with Section 4-4 Diagnostics
Mode on page 4-8.