3.3LSM-503S 
1Specifications 
 
MODEL  
LSM-503S 
Order No. 
 
544-536 
Applicable display unit 
 
LSM-6200 
Laser scanning range 
mm(inch) 
Up to 34 (1.3
”) 
Measuring range 
mm(inch) 
0.3 to 30 (.012
” to 1.18”) 
Resolution  
µm(inch) 
0.02 to 100 (.000001
” to .005”) [Selectable] 
Repeatability [*1] 
µm(inch) 
±0.11 (±.0000044
”)[*2] 
Whole range 
µm(inch) 
±1.0 (±.00004
”)[*3] 
µm 
±( 0.6 + 0.1 
ΔD )   
Linearity  
[*1] 
Narrow 
measuring range 
inch 
±( .000024
” + .000004” 
ΔD ) 
[*3][*4] 
Positional error[*1][*5] 
µm(inch) 
±1.5(±.00006
”) 
Measuring region 
mm(inch) 
10 × 30  (.4
” × 1.18”) 
[Optical axis direction ×
 Scanning direction] 
Number of scans for averaging 
scan 
1to2048 
Laser classification 
 
Class 2 (Max. Output:1.3mW with a scanning laser, semiconductor 
laser: wavelength 650nm) 
Number of laser scans 
/sec 
3200 
Laser scanning rate 
m/sec(inch/sec) 
226 (8898
”/sec) 
Protection level 
 
IP64 
Distance between the laser 
emission unit and reception unit 
mm(inch) 
Standard 130(5.12
”) 
Max. 350 (13
”) [*6] 
Temperature 
°C to 40 °C 
Humidity 
35 %RH to 85 %RH [without condensation] 
Operation environment 
Altitude 
2000 m or less 
Temperature 
-15 
°C to 55 °C 
Storage environment 
Humidity 
35 %RH to 85 %RH [without condensation] 
[*1]:  Environment for accuracy validation: 20
°C 
± 1
°C
 temperature; 50% ± 10% humidity. 
[*2]:  A value of ±2
σ with a 30mm-diameter gage has been measured for two minutes with a measurement interval of 
0.32 seconds, where 
σ is the standard deviation. 
[*3]:  The value of measurements in the center of the measurement region. 
[*4]: 
ΔD is the difference in diameter of the workpiece and the master gage. 
[*5]:  Error due to the positional shift of workpiece in optical axis direction or scanning direction.   
[*6]:  The distance between the laser emission unit and reception unit other than the standard, may affect the accuracy . 
 
2Standard accessories 
 
Part No. 
Item 
Qt. 
― 
ID unit [*7] 
02AGN770A 
Signal cable (5m / 16ft) 
99MBC094A User’s 
manual 
 
[*7]:  The ID unit stores data unique to the measuring unit. A replacement ID unit is available at extra cost.
 
 
3Optional accessories 
 
Order No. 
Item 
02AGD130 
Calibration gages set 
 φ 1 mm   φ 30 mm 
02AGD240 
Air-blow unit for LSM-503 (also available for LSM-503) 
02AGD270 
Workstage for LSM-501 / 503 / 902 (also available for LSM-503) 
02AGD490 
Adjustable Workstage for L SM-503 (also available for LSM-503) 
02AGN780A/B/D 
Extension signal cables ( 5m/10m/20m)(16ft/32ft/64ft)[*8][*10][*11] 
02AGD150A/B 
Extension connecting cables(1m/3m)(3.2ft/9.6ft)[*9][*10][*11] 
[*8]  :  The signal cable can be extended up to 30 m(96ft). 
[*9]  :  The connecting cable can be extended up to 5 m(16ft). 
[*10]:  The total length of the signal cable and the intermediate cable can be extended up to 32 m. 
[*11]:  The length of the 
signal cable and the 
connecting cable other than the standard, may affect the accuracy . 
No. 99MBC094A
 
3-6