mitutoyo 506s. 512s Manuel D’Utilisation

Page de 48
 
3.1LSM-500S 
1Specifications 
 
MODEL 
 LSM-500S 
Order No. 
 544-532 
Applicable display unit 
 LSM-6200 
Laser scanning range 
mm(inch) 
Up to 12.5(.49”)   
(Detecting regions is limited to about 10 mm (.4”) approx.) 
Measuring range 
mm(inch) 
0.005 to 2 (.0002” to .08”) 
0.1 to 2 (.004” to .08”)[*1] 
Resolution  
µm(inch) 
0.01 to 10 (.000001” to .0005”) [Selectable] 
Repeatability[*2] µm(inch) 
±0.03 (±.0000012
”) [*3] 
Linearity [*2] 
µm(inch) 
±0.3 (±.000012
”) [*4] 
Positional error [*2][*5] 
µm(inch) 
±0.4 (±.000016
”) 
Measuring region 
mm(inch) 
1 × 2 (.04
 × .08”) 
[Optical axis direction × Scanning direction] 
Number of scans for averaging 
scan 
16 to 2048[*6] 
Laser classification 
 
Class 2 (Max. Output:1.3mW with a scanning laser, semiconductor laser: 
wavelength 650nm) 
Number of laser scans 
/sec 
3200 
Laser scanning rate 
m/sec (inch/sec) 
76(2992
”/sec) 
Protection level 
 
IP64 
Temperature 0 
°C to 40 °C 
Humidity 
35 %RH to 85 %RH [without condensation] 
Operation environment 
Altitude 
2000 m or less 
Temperature -15 
°C to 55 °C 
Storage environment 
Humidity 
35 %RH to 85 %RH [without condensation] 
[*1]:Measuring range available when set to “No extra-fine wire measurement” or “Edge specification” in the basic setup 
mode. 
[*2]:Environment for accuracy validation: 20
°C
 ± 1
°C
 temperature; 50% ± 10% humidity. 
[*3]:The value of ±2
σ with a 2mm-diameter gage has been measured for two minutes with a measurement interval of 
0.32 seconds, where 
σ is the standard deviation. 
[*4]:The value of measurements in the center of the measurement region. 
[*5]:Error due to the positional shift of the workpiece in the optical axis direction or scanning direction. 
[*6]:Averaging scans between 1 and 8 times can be made if “No extra-fine wire measurement” is specified in the basic 
setup mode. The measuring range, however, is limited to 0.1 mm to 2 mm in this case. 
 
2Standard accessories 
 
Part No. 
Item 
Qt. 
― 
ID unit [*7] 
02AGN770A 
Signal cable (5m / 16ft) 
99MBC094A User’s 
manual 
 
[*7]:The ID unit stores data unique to the measuring unit. A replacement ID unit is available at extra cost. 
 
3Optional accessories 
 
Order No. 
Item 
02AGD110 
Calibration gages set 
 φ 0.1 mm   φ 2 mm 
02AGD200 
Guide pulley unit for LSM-500H (also available for LSM-500) 
02AGD220 
Air-blow unit for LSM-500H (also available for LSM-500) 
02AGN780A/B 
Extension signal cables ( 5m/10m)(16ft/32ft)[*8] [*9] 
[*8]:
The signal cable can be extended up to 20 m(64ft). 
[*9]:The length of the 
signal cable
 other than the standard, may affect the accuracy . 
No. 99MBC094A
 
3-2