IRREGULAR HEARTBEAT SYMBOL (
)
When the monitor detects an irregular rhythm two or more times during
the measurement, the Irregular Heartbeat Symbol (
) will appear on the
display with the measurement values.
An irregular heartbeat rhythm is defined as
a rhythm that is more than 25% slower or
25% faster from the average rhythm
detected while the monitor is measuring the
systolic blood pressure and the diastolic
blood pressure.
If the Irregular Heartbeat Symbol (
)
displays with your measurement results, we
recommend you consult your physician.
Follow the directions of your doctor.
MOVEMENT ERROR SYMBOL (
)
The Movement Error Symbol 
(
is displayed if you move your body
during the measurement. Please remove the arm cuff, and wait 2-3
minutes. Take another measurement.