SPECIFICATIONS
Range
0.00 to 2.00 mg/L
Resolution
0.01 mg/L
Accuracy
±0.03 mg/L ±3% of reading
Typical EMC
±0.01 mg/L
Deviation
Light Source
Tungsten lamp with narrow band interference
filter @ 610 nm
Method
Adaptation of the ASTM Manual of Water and
Environmental Technology,
Environmental Technology,
D859, Heteropoly Blue
method. The reaction between silica and reagents
causes a blue tint in the sample.
REQUIRED REAGENTS
Code
Description
Quantity
HI 93705A-0
Molybdate
6 drops
HI 93705B-0
Citric acid
1 packet
HI 93705C-0
Amino acid
1 packet
REAGENT SETS
HI 93705-01 Reagents for 100 tests
HI 93705-03 Reagents for 300 tests
For other accessories see page 141.
HI 93705-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Silica on the
secondary LCD by pressing PROGRAM W and V.
• Fill the cuvet with 10 mL of unreacted sample (up to
the mark).
• Add 6 drops of HI 93705A Molybdate
reagent. Replace the cap and swirl
the solution.
the solution.
• Wait for 4 minutes, add the content
of one packet of HI 93705B Citric
acid reagent and shake until it is
completely dissolved.
acid reagent and shake until it is
completely dissolved.
Silica
10 mL
x 6