Hanna Instruments c 99 Manuel D’Utilisation

Page de 148
131
SPECIFICATIONS
Range
0.000 to 1.000 mg/L
Resolution
0.001 mg/L
Accuracy
±0.005 mg/L ±10% of reading
Typical EMC
± 0.001 mg/L
Deviation
Light Source
Tungsten lamp with narrow band interference
filter @ 575 nm.
Method
Adaptation of the PAN method. The reaction
between silver and reagents causes an orange
tint in the sample.
REQUIRED REAGENTS
Code
Description
Quantity
HI 93737A-0
Buffer Reagent A
1 mL
HI 93737B-0
Buffer Reagent B
1 mL
HI 93737C-0
Indicator Reagent C 2 mL
HI 93737D-0
Fixing Reagent D
2 mL
HI 93703-51
Dispersing Agent
4-6 drops
(only when necessary, see note)
REAGENT SETS
HI 93737-01 Reagents for 50 tests
HI 93737-03 Reagents for 150 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Silver on the
secondary LCD by pressing PROGRAM W and V.
Note: for best results perform your tests between 20-24°C.
• Fill two graduated beakers with 25 mL of sample.
• Add 1.0 mL of HI 93737A reagent
to only one beaker (the blank) and
swirl gently to mix.
SILVER
SILVER
SILVER
SILVER
SILVER
Silver
# 1
# 2
# 1