Hanna Instruments c 99 Manuel D’Utilisation

Page de 148
21
A L U M I N U M
A L U M I N U M
A L U M I N U M
A L U M I N U M
A L U M I N U M
Aluminum
SPECIFICATIONS
Range
0.00 to 1.00 mg/L
Resolution
0.01 mg/L
Accuracy
±0.02 mg/L ±4% of reading
Typical EMC
±0.01 mg/L
Deviation
Light Source
Tungsten lamp with narrow band interference
filter @ 525 nm
Method
Adaptation of the aluminon method. The reac-
tion between aluminum and reagents causes a
reddish tint in the sample.
REQUIRED REAGENTS
Code
Description
Quantity
HI 93712A-0
Ascorbic acid
1 packet
HI 93712B-0
Aluminon reagent
1 packet
HI 93712C-0
Bleaching powder
1 packet
REAGENT SETS
HI 93712-01 Reagents for 100 tests
HI 93712-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Aluminum on the
secondary LCD by pressing PROGRAM W and V.
• Fill a graduated beaker with 50 mL of sample.
• Add the content of one packet of HI 93712A
Ascorbic acid reagent and mix until dissolution is
complete.
• Add the content of one packet of HI 93712B
Aluminon reagent and mix until dissolution is
complete. This is the sample.