Hanna Instruments c 99 Manuel D’Utilisation

Page de 148
31
Free Chlorine
10 mL
FREE CHLORINE
FREE CHLORINE
FREE CHLORINE
FREE CHLORINE
FREE CHLORINE
SPECIFICATIONS
Range
0.00 to 2.50 mg/L
Resolution
0.01 mg/L
Accuracy
±0.03 mg/L ±3% of reading
Typical EMC
±0.01 mg/L
Deviation
Light Source
Tungsten lamp with narrow band interference
filter @ 525 nm
Method
Adaptation of the EPA DPD method 330.5. The
reaction between free chlorine and the DPD re-
agent causes a pink tint in the sample.
REQUIRED REAGENTS
POWDER
:
Code
Description
Quantity
HI  93701-0
DPD
1 packet
LIQUID:
Code
Description
Quantity
HI  93701A-F
DPD1 Indicator
3 drops
HI  93701B-F
DPD1 Buffer
3 drops
REAGENT SETS
HI 93701-F Reagents for 300 tests (liquid)
HI 93701-01 Reagents for 100 tests (powder)
HI 93701-03 Reagents for 300 tests (powder)
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Free Chlorine on the
secondary LCD by pressing PROGRAM W and V.
• Fill the cuvet with 10 mL of unreacted sample (up
to the mark) and replace the cap.
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.