Hanna Instruments c 99 Manuel D’Utilisation

Page de 148
48
Chromium VI LR
SPECIFICATIONS
Range
0 to 300 µg/L
Resolution
1 µg/L
Accuracy
±1 µg/L ±4% of reading
Typical EMC
±1 µg/L
Deviation
Light Source
Tungsten lamp with narrow band interference
filter @ 525 nm
Method
Adaptation of the ASTM Manual of Water and
Environmental Technology, D1687-92
,
Diphenylcarbohydrazide method. The reaction be-
tween chromium VI and the reagent causes a
purple tint in the sample.
REQUIRED REAGENTS
Code
Description
Quantity
HI 93749-0
Powder  reagent
1 packet
REAGENT SETS
HI 93749-01 Reagents for 100 tests
HI 93749-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Chromium VI LR on
the secondary LCD by pressing PROGRAM W and V.
• Fill the cuvet with 10 mL of unreacted sample
(up to the mark) and replace the cap.
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Press ZERO and “SIP” will blink on the display.
• Wait for a few seconds and the display will
show “-0.0-”. Now the meter is zeroed
and ready for measurement.
CHROMIUM VI LOW RANGE
CHROMIUM VI LOW RANGE
CHROMIUM VI LOW RANGE
CHROMIUM VI LOW RANGE
CHROMIUM VI LOW RANGE
10 mL