Hanna Instruments c 99 Manuel D’Utilisation

Page de 148
56
CYANIDE
CYANIDE
CYANIDE
CYANIDE
CYANIDE
Cyanide
SPECIFICATIONS
Range
0.000 to 0.200 mg/L
Resolution
0.001 mg/L
Accuracy
  ±0.005 mg/L ±3% of reading
Typical EMC Dev. ±0.001 mg/L
Light Source
Tungsten lamp with narrow band interference
filter @ 610 nm
Method
Adaptation of the Standard Methods for the
Examination of Water and Wastewater, 18
th
edition
, Pyridine-Pyrazolone method. The reac-
tion between cyanide and reagents causes a
blue tint in the sample.
REQUIRED REAGENTS
Code
Description
Quantity
HI 93714A-0
Reagent A
1 spoon
HI 93714B-0
Reagent B
1 packet
HI 93714C-0
Reagent C
1 packet
REAGENT SETS
HI 93714-01 Reagents for 100 tests
HI 93714-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Cyanide on the
secondary LCD by pressing PROGRAM W and V.
• Fill the cuvet with 10 mL of unreacted sample (up
to the mark) and replace the cap.
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Press ZERO and “SIP” will blink on the display.
• Wait for a few seconds and the display will
show “-0.0-”. Now the meter is zeroed
and ready for measurement.
10 mL