Hanna Instruments c 99 Manuel D’Utilisation
64
CALCIUM HARDNESS
CALCIUM HARDNESS
CALCIUM HARDNESS
CALCIUM HARDNESS
CALCIUM HARDNESS
SPECIFICATIONS
Range
0.00 to 2.70 mg/L
Resolution
0.01 mg/L
Accuracy
±0.11 mg/L ±5% of reading
Typical EMC
±0.01 mg/L
Deviation
Light Source
Tungsten lamp with narrow band interference
filter @ 525 nm
Method
Adaptation of the Standard Methods for the
Examination of Water and Wastewater, 18
Examination of Water and Wastewater, 18
th
edition,
Calmagite method. The reaction between
calcium and reagents causes a reddish-violet tint
in the sample.
REQUIRED REAGENTS
Code
Description
Quantity
HI 93720A-0
Ca & Mg indicator
0.5 mL
HI 93720B-0
Alkali solution
0.5 mL
HI 93720C-0
EGTA solution
1 drop
REAGENT SETS
HI 93720-01 Reagents for 100 tests
HI 93720-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Hardness Ca on the
secondary LCD by pressing PROGRAM W and V.
• Rinse a graduated beaker
several times with unreacted
sample, before filling it to
the 50 mL mark with the
sample.
• Add 0.5 mL of HI 93720A Calcium
indicator solution and swirl to mix.
• Add 0.5 mL of
HI 93720B Alkali solu-
tion and swirl to mix.
Use this solution to rinse
2 cuvets before filling
them up to the 10 mL
mark.
Hardness Ca
# 1
# 2