Hanna Instruments c 99 Manuel D’Utilisation

Page de 148
64
CALCIUM HARDNESS
CALCIUM HARDNESS
CALCIUM HARDNESS
CALCIUM HARDNESS
CALCIUM HARDNESS
SPECIFICATIONS
Range
0.00 to 2.70 mg/L
Resolution
0.01 mg/L
Accuracy
±0.11 mg/L ±5% of reading
Typical EMC
±0.01 mg/L
Deviation
Light Source
Tungsten lamp with narrow band interference
filter @ 525 nm
Method
Adaptation of the Standard Methods for the
Examination of Water and Wastewater, 18
th
edition, 
Calmagite method. The reaction between
calcium and reagents causes a reddish-violet tint
in the sample.
REQUIRED REAGENTS
Code
Description
Quantity
HI 93720A-0
Ca & Mg indicator
0.5 mL
HI 93720B-0
Alkali solution
0.5 mL
HI 93720C-0
EGTA solution
1 drop
REAGENT SETS
HI 93720-01 Reagents for 100 tests
HI 93720-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Hardness Ca on the
secondary LCD by pressing PROGRAM W and V.
• Rinse a graduated beaker
several times with unreacted
sample, before filling it to
the 50 mL mark with the
sample.
• Add 0.5 mL of HI 93720A Calcium
indicator solution and swirl to mix.
• Add 0.5 mL of
HI 93720B Alkali solu-
tion   and swirl to mix.
Use this solution to rinse
2 cuvets before filling
them up to the 10 mL
mark.
Hardness Ca
# 1
# 2