Hanna Instruments c 99 Manuel D’Utilisation
67
MAGNESIUM HARDNESS
MAGNESIUM HARDNESS
MAGNESIUM HARDNESS
MAGNESIUM HARDNESS
MAGNESIUM HARDNESS
Hardness Mg
SPECIFICATIONS
Range
0.00 to 2.00 mg/L
Resolution
0.01 mg/L
Accuracy
±0.11 mg/L ±5% of reading
Typical EMC
±0.02 mg/L
Deviation
Light Source
Tungsten lamp with narrow band interference
filter @ 525 nm
Method
Adaptation of the Standard Methods for the
Examination of Water and Wastewater, 18
Examination of Water and Wastewater, 18
th
edition,
EDTA colorimetric method. The reaction
between magnesium and reagents causes a red-
dish-violet tint in the sample.
REQUIRED REAGENTS
Code
Description
Quantity
HI 93719A-0
Mg indicator
0.5 mL
HI 93719B-0
Alkali solution
0.5 mL
HI 93719C-0
EDTA solution
1 drop
HI 93719D-0
EGTA solution
1 drop
REAGENT SETS
HI 93719-01 Reagents for 100 tests
HI 93719-03 Reagents for 300 tests
For other accessories see page 141.
HI 93719-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Hardness Mg on the
secondary LCD by pressing PROGRAM W and V.
• Rinse a graduated beaker several times
with unreacted sample, before filling it
to the 50 mL mark with the sample.
to the 50 mL mark with the sample.
• Add 0.5 mL of HI 93719A Magnesium
indicator solution, then swirl to mix.