National Instruments NI 783xR Manuel D’Utilisation

Page de 73
Chapter 2
Hardware Overview of the NI 783xR
2-20
ni.com
delay-matched trigger signal between the first peripheral slot and the 
other peripheral slots for precise trigger timing signals. For example, an 
NI PXI-783xR in Slot 2 can send an independent trigger signal to each 
device plugged into Slots <3..15> using the PXI/LBLSTAR<0..12>. Each 
device receives its trigger signal on its own dedicated star trigger line.
Caution
Do not configure the NI 783xR and another device to drive the same physical star 
trigger line simultaneously. Such signal driving can damage the NI 783xR and the other 
device. NI is not liable for any damage resulting from such signal driving.
Refer to the PXI Hardware Specification Revision 2.1 and PXI Software 
Specification Revision 2.1
 at 
www.pxisa.org
 for more information about 
PXI triggers.
Switch Settings
Refer to Figure 2-11 for the location of switches on the NI PXI-783xR and 
Figure 2-12 for the location of switches on the NI PCI-783xR. For normal 
operation, SW1 is in the OFF position. To prevent a VI stored in Flash 
memory from loading to the FPGA at power up, move SW1 to the ON 
position, as shown in Figure 2-13. SW2 and SW3 are not connected.