National Instruments SCXI-1121 Manuel D’Utilisation

Page de 162
Chapter 2
Configuration and Installation
2-10
www.natinst.com
Excitation Level
Each excitation channel of your SCXI-1121 has two different current or 
voltage levels. You can set a given channel to one of the following level 
modes:
In the Current mode
0.150 or 0.450 mA
In the Voltage mode
3.333 or 10 V
It is important to notice that you should select the level of excitation 
according to the load you are using. Table 2-7 lists the maximum load that 
can be driven per channel at each level of excitation for both volt and 
current excitation.
Table 2-6.  Voltage and Current Mode Excitation Jumper Setup
Excitation 
Channel
Jumpers
Voltage Mode
(Factory Setting)
Current Mode
0
W14 and W15
1
W22 and W23
2
W34 and W35
3
W46 and W47
W14
1    2    3
W15
1    2    3
W14
1    2    3
W15
1    2    3
W22
1    2    3
W23
1    2    3
W22
1    2    3
W23
1    2    3
W34
1    2    3
W35
1    2    3
W34
1    2    3
W35
1    2    3
W46
1    2    3
W47
1    2    3
W46
1    2    3
W47
1    2    3