Chapter 2
Configuration and Installation
© National Instruments Corporation
2-15
Signal Connections
This section describes the input and output signal connections to the 
SCXI-1121 board via the SCXI-1121 front connector and rear signal 
connector, and includes specifications and connection instructions for the 
signals given on the SCXI-1121 connectors.
Cautions
Do not operate the device in an explosive atmosphere or where there may be 
flammable gasses or fumes.
Keep away from live circuits. Do not remove equipment covers or shields unless you are 
trained to do so. If signal wires are connected to the device, hazardous voltages may exist 
even when the equipment is turned off. To avoid a shock hazard, do not perform procedures 
involving cover or shield removal unless you are qualified to do so and disconnect all field 
power prior to removing covers or shields.
Equipment described in this document must be used in an Installation Category II 
environment per IEC 60664. This category requires local level supply mains-connected 
installation.
Do not operate damaged equipment. The safety protection features built into this device 
can become impaired if the device becomes damaged in any way. If the device is damaged, 
turn the device off and do not use until service-trained personnel can check its safety. If 
necessary, return the device to National Instruments for service and repair to ensure that its 
safety is not compromised.
Do not operate this equipment in a manner that contradicts the information specified in this 
document. Misuse of this equipment could result in a shock hazard.
Terminals are for use only with equipment that has no accessible live parts.
Do not substitute parts or modify equipment. Because of the danger of introducing 
additional hazards, do not install unauthorized parts or modify the device. Return the 
device to National Instruments for service and repair to ensure that its safety features are 
not compromised.
When using the device with high common-mode voltages, you must insulate your signal 
wires for the highest input voltage. National Instruments is not liable for any damages or 
injuries resulting from inadequate signal wire insulation. Use only 26-14 AWG wire with 
a voltage rating of 300 V and 60 °C for measuring 250 V. Prepare your signal wire by 
stripping the insulation no more than 7 mm.