eutech-instruments cyberscan con 11 conductivitytds meter Manuel D’Utilisation

Page de 80
Instruction Manual 
 
CON 11/110 
 
13
4.3 Temperature 
Calibration 
Your electrode features a built-in temperature sensor. The temperature sensor is 
factory calibrated. Calibrate your sensor only if you suspect temperature errors may 
have occurred over a long period of time or if you have a 
replacement electrode. 
1.  Make sure the electrode is attached to the 6-pin 
connector. The ATC annunciator will appear at the 
bottom right-hand corner of the LCD.  
2.  Switch the meter on. Press the CAL/MEAS key to 
enter conductivity or TDS calibration mode. The CAL 
indicator appears above the primary display.  
3.  While in conductivity or TDS calibration mode, press 
the MODE key to enter temperature calibration 
mode. The primary display shows the current 
measured temperature reading based on the last set 
offset and the secondary display shows the current 
measured temperature reading based on factory 
default calibration.  
4.  Dip the electrode into a solution of known 
temperature (i.e. a temperature bath). Allow time for 
the built-in temperature sensor to stabilise.  
5. 
Scroll with the MI/▲ or MR/▼ keys to set the correct 
temperature value (i.e. the temperature of the 
temperature bath). You can adjust the reading in 
increments of 0.1 °C.  
6.  Once you have selected the correct temperature, 
press the HOLD/ENTER key to confirm. The meter 
will be calibrated and return to the measurement 
mode.  
NOTES:  
•  To exit this program without confirming the 
temperature calibration value, DO NOT press 
HOLD/ENTER. Press CAL/MEAS instead.  
•  If the ATC indicator does not light up, see SETUP 
menu Program P1.3 to switch it on. See page 40. 
•  Since temperature readings affect the accuracy of 
conductivity / TDS measurements, it is strongly 
recommended to carry out a conductivity / TDS 
calibration after a temperature calibration is done. 
 
Figure 11: Temperature 
Calibration
 
MEAS
1413
1413
ATC
°C
22.3
1413
MEAS
ATC
°C
22.0
CAL
CAL
ATC
°C
22.3
22.3
ATC
°C
22.0
22.3
CAL
READY
µS
µS
1425
µS
MODE
  CAL
MEAS
HOLD
ENTER
MI
MR