eutech-instruments cyberscan con 11 conductivitytds meter Manuel D’Utilisation

Page de 80
Instruction Manual 
 
CON 11/110 
 
17
4.6 
Manual Calibration (For Conductivity & TDS Calibration) 
In Manual Calibration mode, you can use customised Conductivity / TDS calibration 
standards (specific to your own application) and calibrate the meter. The following 
example shows the calibration sequence to 12.00 mS conductivity calibration 
standard. Set up the manual calibration mode in the SETUP menu Program P9.1 by 
setting to “NO”. See page 51. 
The following procedure describes the method for a single point calibration. Set the 
single point calibration mode to on in the SETUP menu Program 9.2 by setting to 
‘YES”. See page 52 for the single point calibration set up procedure. 
1.   If necessary, press the MODE key to select 
conductivity mode.   
2. Rinse 
the 
electrode 
thoroughly with de-ionised 
water or a rinse solution, and then rinse with a 
small amount of calibration standard. 
3.  Dip the electrode into the calibration standard. 
Immerse the electrode tip beyond the upper steel 
band (see Figure 6 in page 6). Stir the electrode 
gently to create a homogeneous sample. Allow 
time for the reading to stabilise. 
4.  Press CAL/MEAS key to enter conductivity 
calibration mode. The CAL indicator will appear 
in the upper corner of the display. The upper 
(primary) display shows the measured value 
which is with respect to the last calibration while 
the lower (secondary) display shows the 
measured value with respect to the default 
calibration. 
NOTE: To exit calibration without confirmation, press 
CAL/MEAS key again to go back to measurement 
mode.  
5.  Wait for the value to stabilise and press MI/▲ or 
MR/▼ key and adjust the value in the upper 
display to the calibration standard used.  
6.  Press the HOLD/ENTER key to confirm. The 
meter returns to measurement mode with the 
new calibrated value.  
This single point calibration will apply to all the five ranges. 
As for multi point calibration, set up the multi point 
calibration mode in the SETUP menu Program P9.2 by setting to ‘NO”.  
Repeat step 1 to 6 for every calibration point (one in each measuring range) using 
the desired calibration solutions until all points have been calibrated. 