5.2.1  Setting the manual temperature compensation value 
To use manual temperature compensation (MTC), you need to enter the temperature 
value of your sample into the meter. This is the temperature value at which the 
measured readings will be manually temperature compensated. You can select any 
temperature between -10.0 to 110.0 °C (14.0 to 230 °F). Default value is 25 °C.  
1.  Switch the meter on. If necessary, set up the MTC by selecting ATC off as 
described in page 21. The ATC indicator will not appear on the display.  
2.  Press the CAL/MEAS key to enter conductivity or 
TDS calibration mode. The CAL indicator will appear 
above the primary display. 
3.  While in conductivity or TDS calibration mode, press 
the MODE key to enter temperature calibration 
mode. The primary display act as the adjustable 
temperature setting and the secondary display 
shows the default temperature value of 25 °C or 
indicates the last set temperature setting. Check the 
temperature of your sample using an accurate 
thermometer. 
4.  Press the MI/▲ or MR/▼ keys to offset the 
temperature to the measured value from step 3.  
5.  Press HOLD/ENTER to confirm the selected 
temperature and to return to the Conductivity or TDS 
measurement mode.  
The meter will now compensate Conductivity or TDS 
readings for manually set temperature.  
NOTES: 
To exit this program without confirming the manual 
temperature compensation value, DO NOT press 
HOLD/ENTER in step 5. Press CAL/MEAS instead.  