CON 11/110 
 
36
8.1 
Advanced SETUP Mode Overview  
1.  Press the SETUP key to enter Set Up menu. 
2. Press 
the 
MI/▲ or MR/▼ keys to scroll through sub groups.  
3.  Press HOLD/ENTER key to enter a particular parameter.  
SETUP
 P 1.0
COF
 
P1.0: Unit Configuration 
P1.1  Selection of READY and *Auto HOLD function (*CON 
 110 
Only). 
P1.2  Selection of °C or *°F (*CON 110 Only) 
P1.3  Selection of Automatic or Manual Temperature 
 Compensation 
P1.4  Selection of TDS Conversion Factor (This mode only 
 
appears when SETUP Menu is chosen while in the 
 
TDS measurement mode) 
SETUP
CAL
 P 2.0
 
P2.0: Viewing previous calibration data 
P2.1  First range calibration point 
P2.2  Second range calibration point 
P2.3  Third range calibration point 
P2.4  Fourth range calibration point 
P2.5  Fifth range calibration point 
SETUP
ELE
 P 3.0
 
P3.0: Viewing electrode data 
P3.1  Effective cell constant for first range 
P3.2  Effective cell constant for second range  
P3.3  Effective cell constant for third range 
P3.4  Effective cell constant for fourth range 
P3.5  Effective cell constant for fifth range 
SETUP
 P 4.0
A o
 
P4.0: Auto Off 
P4.0 Selection 
of 
automatic power off function  
SETUP
 P 5.0
 
P5.0: Reset to factory defaults 
P5.1  Resetting of all calibration and clear all 
 calibration 
data 
P5.2  Resetting of all User settings to factory 
 defaults 
condition 
SETUP
 P 6.0
MEM
 
P6.0: Memory clear 
P6.0   Clearance of all stored readings 
SETUP
 P 7.0
°C °F
 
P7.0: Temperature  
P7.1  Setting of temperature coefficient 
P7.2  Setting of normalisation temperature