eutech-instruments cyberscan con 11 conductivitytds meter Manuel D’Utilisation

Page de 80
Instruction Manual 
 
CON 11/110 
 
45
8.6 
P5.0: Reset to factory defaults  
Program 5.0 reset functions lets you have an option to only reset all the meter’s 
calibrations and its data that have been stored or reset all calibrations, its stored 
data, memory, and any other setup functions that you might have changed back to 
factory default settings. See Factory Default Settings  at page 69 and 70. 
8.6.1  P5.1: Calibration Reset  
This reset function clears all the calibrations and its data that 
have been set by you. All other setup parameters that you 
have changed are not reset. 
From measurement mode 
1.  Press SETUP key to enter Set Up menu.  
2.  Press the MI/▲ or MR/▼ keys to until you view 
parameter P5.0.  
3.  Press the HOLD/ENTER key to enter parameter 
P5.1  with the upper display showing “CAL”. 
4.  Press the HOLD/ENTER key again to enter the 
calibration reset selection mode.  
5.  Press the MI/▲ or MR/▼ keys to toggle between NO 
and YES. 
• 
NO retains current settings 
• 
YES clears all calibrations and its data. 
6.  Press the HOLD/ENTER key to confirm selection 
and to return to measurement mode. Otherwise 
press CAL/MEAS key to return to measurement 
mode without resetting the calibration data. 
 
 
SETUP
cal
p 5.1
SETUP
cal
SETUP
SETUP
yes
MI
MR
HOLD
ENTER
HOLD
ENTER
p 5.0
cal
Figure 38: P5.1-Clear all 
calibration and its data