eutech-instruments cyberscan con 11 conductivitytds meter Manuel D’Utilisation

Page de 80
Instruction Manual 
 
CON 11/110 
 
67
12 TROUBLE 
SHOOTING 
 
Problem Cause 
Solution 
No display when 
‘ON’ key is pressed 
a)  Batteries not in place 
b)  Batteries not in correct polarity (+ 
and – position) 
c) Weak 
batteries 
d)  No AC power to adapter (if this 
optional accessory is used) 
a)  Check that batteries are in 
place and making good 
contact 
b)  Re-insert batteries with correct 
polarity 
c) Replace 
batteries 
d)  Check power to AC adapter 
Unstable readings 
a)  Air bubbles in electrode. 
b) Dirty electrode 
c)  Electrode not deep enough in 
sample 
d)  External noise pickup or induction 
caused by nearby electric motor 
e)  Electrode guard not attached 
 
a)  Tap electrode to remove 
bubbles, move electrode away 
from bubbles is present in 
sample 
b)  Clean the electrode and re-
calibrate  
c)  Make sure sample entirely 
covers the electrode sensors 
d)   Move or switch off interfering 
motor.  
e)  Attach electrode guard 
Slow response 
a)  Dirty / Oily electrode  
a)  Clean electrode. See 
“Electrode Care & 
Maintenance”, page 65 
Meter is not 
responding to key 
press. 
a)  HOLD mode in operation 
 
a)  Press the HOLD/ENTER key 
to cancel the HOLD mode 
ATC indicator blinks 
and temperature 
display shows “Ur” 
a)  Improper attachment of electrode 
to the meter 
b)  electrode built-in temperature 
sensor cable connection has 
broken 
a)  Check that the electrode 
connects properly to the meter 
b) Replace electrode