7.6 Current Output (rng) sub-function
This sub-function lets you set the transmitter current output range of this unit. The difference between
the upper and lower range has to be a minimum of 20% of Full Scale, anywhere on the scale.
7.6.1 Entering current output sub-function
1) Enter Advanced Set-up mode. Push the ENTER key and scroll
to Advanced Set-up security code “22”. Push the ENTER key again.
2) Press the 
∆∆
 or 
 keys to scroll until the upper display shows
“rng”. Press ENTER.
3) Measurement range selection. Please refer to 7.7. Press
ENTER to proceed to current output sub-function (7.8.2)
7.6.2 Choosing the output type
This parameter lets you choose between 0-20 mA or 4-20 mA output.
1) Follow directions in 7.8.1 to enter Current Output mode.
2) Press the ENTER key. Scroll with the 
∆∆
 or 
 keys until the upper
display shows the output type (0-20 or 4-20), and the lower display
shows “out”.
3) Press the 
∆∆
 or 
 keys to select your output type: 0-20 or 4-20
mA.
4) Press the ENTER key to confirm your selection.
5) Proceed to 7.8.3, or return to Measurement mode by pressing
the 
∆∆
 and 
 keys simultaneously (escape).
7.6.3 Selecting Conductivity value at 0(4)mA
This parameter lets you choose the conductivity value at which the
transmitter output will be 0(4) mA.
Follow directions in 7.8.2 to enter Current Output mode.
1) Press the ENTER key. Scroll with the 
∆∆
 or 
 keys until the upper display shows a conductivity
value and the lower display shows “r.0(4)”.
2) Press the 
∆∆
 or 
 keys to select the required conductivity value to be equivalent to 0(4) mA (0.00
to 80.0 % F.S.; Default is 0% F.S.).
3) Press the ENTER key to confirm your selection.
4) Press the ENTER key to return to Advanced Set-up mode, or return to Measurement mode by
pressing the 
∆∆
 and 
 keys simultaneously (escape).
7.6.4 Selecting Conductivity value at 20mA
This parameter lets you choose the conductivity value at which the transmitter output will be 20mA.
5) Follow directions in 7.8.3 to enter Current Output mode.
6) Press the ENTER key. Scroll with the 
∆∆
 or 
 keys until the upper display shows a conductivity
value and the lower display shows “r.20”.
7) Press the 
∆∆
 or 
 keys to select the required conductivity value to be equivalent to 20 mA (20.0 to
100.0% F.S.; Default is 100% F.S.).