eutech-instruments ecoscan do 6 do meter Manuel D’Utilisation
Instruction Manual
DO 6
18
blink and the display will show a value which is with respect to the factory
default calibration.
default calibration.
NOTES:
When calibration is attempted on a factory calibrated absolute value
measurement of 10.1% to 49.9%, the display will show the absolute value
measurement for 1.5 seconds before displaying Err.1 and goes to the
measurement mode.
measurement for 1.5 seconds before displaying Err.1 and goes to the
measurement mode.
5) Press the HOLD/ENTER key to confirm the calibration. The meter displays
“CO” for 1.5 seconds and automatically calibrates to 100.0% air saturation
and returns to measurement mode.
and returns to measurement mode.
Note: You can offset your % DO calibration. See Section 7.7 (OFS.6) %
Saturation Offset Adjustment for directions.
Saturation Offset Adjustment for directions.