23
 
5.2 Manual 
Temperature 
Compensation 
  For 
manual 
temperature 
compensation, simply deactivate the 
“A.ATC” mode as explained in the set 
up menu on page 36. The automatic 
temperature compensation annunciator 
“A” will not be displayed to represent 
that the meter is in manual temperature 
compensation mode.  
To use manual temperature 
compensation, you need to enter the 
temperature value of your process into 
the meter. You can select any 
temperature between 0 and 50 °C. 
Default value is 25 °C.  
1)  Switch on the meter and if 
necessary, press MODE key to 
select temperature measurement 
mode.  
2)  Press CAL key to start temperature 
calibration process.  
3)  The “CAL” indicator will start 
blinking and the display will show 
the last manual set temperature 
value.  
4)  Check the temperature of your 
sample using an accurate 
reference thermometer. Wait for 
the thermometer value to stabilize. 
Press the ▲ or ▼ key of your 
meter to manually set its 
temperature value according to the 
CAL 
CAL 
T
T
T
T
CAL
HOLD
ENTER
Last set  manual  
temperature setting
“CO” will display for 1.5 
seconds
Figure 14: Manual Temperature 
Compensation value selection sequence