7.6 
(rSt.5) Reset to Factory Default  
This mode lets you reset all parameters to 
factory default settings. There are two 
levels of reset:  
1)  Calibration Reset: - Reset only the 
calibration values. This clears all 
previous calibrated values and reset 
to factory default. See Section 11 for 
factory default settings. 
2)  User Reset: - Clears all data including 
calibration and other customized 
setup functions and reset it to factory 
default. See Section 11 for factory 
default settings. 
7.6.1 
 Calibration Reset  
From any Set Up menu: 
1)  Scroll the Set Up main group menu 
by using the ▲ and ▼ keys until the 
meter displays the Reset to Factory 
Default menu “rSt.5”. 
2)  Press the HOLD/ENTER key to enter 
the menu. 
3)  Press the HOLD/ENTER key to enter 
the Calibration Reset menu “A.CAL”. 
4) Use 
the 
▲ and ▼ keys to activate or 
deactivate the Calibration Reset 
selection. 
a. YES – activates Calibration 
Reset 
Figure 26: Calibration reset 
selection
 
HOLD
ENTER
HOLD
ENTER
Calibration
 Reset