PROBE CARE AND MAINTENANCE 
The DO6 probe is a galvanic measuring element which produces an output 
proportional to the oxygen present in the medium in which it is placed. The 
galvanic probe design lets you take measurements immediately – without the 
typical 15 minute wait of other dissolved oxygen probes.  
The probe consists of two parts: 
• 
An upper part consisting of the anode, a cathode, and the dual cable.  
• 
A lower part consisting of a pre-membraned cap, and electrolyte 
solution.  
Oxygen diffuses through the membrane onto the cathode, where it is 
consumed. This process produces an electrical current which flows through the 
cable to the meter. The electric current produced is proportional to the oxygen 
that passes through the membrane and the layer of electrolyte.  This makes it 
possible to measure the partial pressure of oxygen in the sample at a given 
temperature.  
Since the DO in the sample is consumed by the cathode it is essential that a 
new sample must flow past the membrane of the probe to prevent the 
occurrence of false readings. The probe uses very little oxygen for its 
measurement. This enables it to function correctly with liquid movement as low 
as 2 inch /sec across membrane. 
The permeability of the membrane to oxygen varies greatly with temperature. 
Therefore compensation is needed for this variation. The  DO6 probe comes 
with an in-built Temperature Compensation for the membrane variation.