eutech-instruments sodium glass Manuel D’Utilisation

Page de 16
Sodium Electrode 
 
Instruction Manual 
 
2
 
6. 
Eutech Sodium Standard, 100 ppm Na
+
, Code no. EC-SCS-LNA3-BT. To prepare this 
solution from your own laboratory stock, half fill a one liter volumetric flask with 
distilled water and add 0.254 grams of reagent grade NaCl. Swirl the flask to dissolve the 
solid. Fill the flask to the mark with distilled water, cap, and upend several times to mix 
the solution. 
 
 
GENERAL PREPARATION 
 
Electrode Preparation 
 
Remove the rubber caps covering the electrode tips and the rubber insert covering the filling hole of 
the reference electrode. Fill the combination electrode or the reference electrode with the filling 
solution shipped with the electrode to a level just below the fill hole. No preparation is required 
with a sealed reference electrode. Connect the electrodes to the proper terminals of the meter as 
recommended by the meter manufacturer. Soak the sodium electrode tip overnight in the electrode 
storage solution. 
 
Electrode Slope Check (with pH/mV meter)   
(Check electrodes each day) 
 
1. 
To a 150 ml glass beaker, add 100 ml of distilled water and 2 ml of ISA. Place the beaker 
on a magnetic stirrer and begin stirring at a constant rate. After assuring that the meter is 
in the millivolt mode, lower the electrode tips into the solution. 
 
2. 
Using a pipet, add 1 ml of 0.1M or 1,000 ppm sodium standard to the beaker. When the 
reading has stabilized, record the mV reading. 
 
3. 
Using a pipet, add 10 ml of the same sodium standard used above to the beaker. When 
the reading has stabilized, record the mV reading. 
 
4. 
Determine the difference between the two readings. A difference of 56
±2 mV indicates 
correct electrode operation assuming the solution temperature is between 20
o
 and 25
o
C.  
See the 
TROUBLESHOOTING
 section if the potential change is not within this range. 
       
Slope is defined as the change in potential observed when the concentration changes be a factor of 
10. 
 
Electrode Slope Check (with ion meter) 
(Check electrodes each day) 
 
1. 
Prepare standard sodium solutions whose concentrations vary by tenfold. Use either the 
0.1M NaCl or the 1,000 ppm sodium standard. Use the serial dilution method for this 
preparation. 
 
2. 
To a 150 ml beaker, add 100 ml of the lower value standard and 2 ml of ISA. Place the 
beaker on the magnetic stirrer and begin stirring at a constant rate. Lower the electrode 
tips into the solution. Assure that the meter is in the concentration mode. 
 
3. 
Adjust the meter to the concentration of the standard and fix the value in the memory 
according to the meter manufacturer's instructions.