3.2. Mounting Hardware 
 
Eutech Instruments’ short nose sensor (Fig 3.2) has insertion threads located closer to the sensing 
elements and is recommended for short branch tees (typically metal tees and SCH 40 plastic tees).  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Save the protective cap and use it to 
keep electrodes moist whenever the 
sensor is removed from service. 
 
When taking a sensor out of storage, 
soak the sensor in a pH 7 buffer 
solution for 30 minutes before 
calibration. 
 
If the sensor does not calibrate, refer 
to Section 4 to review maintenance, 
refurbishment and troubleshooting 
options. 
 
The sensor should be positioned at 
least 15
° up from horizontal.  This 
will insure that entrained air does not 
form a pocket around the sensor 
head. 
 
DO NOT USE PIPE SEALANT.  
Pipe sealants may not provide 
adequate sealing or may react with 
different plastic materials. Use 
thread tape. 
 
DO NOT OVER TIGHTEN! 
Maximum torque: 66.4 lbs
⋅inch (1.13 N⋅m) 
Figure 3.2: Short Nose Sensor Body 
Dimensions: IN (mm)