4.  pH SENSOR MAINTENANCE 
4.1. Cleaning the pH Sensor Head 
 
4.1.1.  In order to maintain an accurate measurement value, the sensor will need occasional maintenance.  
The maintenance interval will be dictated by the process in which it is installed.  The harsher the 
process, the more often the sensor will require maintenance.  Regular maintenance will yield a longer 
sensor life. 
 
4.1.2.  The sensor cleaning procedure is as follows: 
 
4.1.2.1. Remove sensor from service and rinse or spray it with warm water to remove heavy deposits. 
 
4.1.2.2. Soak the sensor in a container of hot detergent water for one hour.  Do not use detergents that 
contain oily skin softeners like aloe or lanolin that can coat the glass electrode.  Alconox
®
 and 
Dawn™ dishwashing liquid work well. 
 
4.1.2.3. Use a soft-bristled brush, such as a soft toothbrush, and hot detergent water to scrub the entire 
electrode end of the sensor, being careful not to scratch or break the glass electrode. 
 
4.1.2.4. Rinse the electrode end with clean warm water. 
 
4.1.2.5. If deposits are still present on glass electrode repeat steps 2 and 3.  In the case of lime or other 
mineral deposits a weak solution (about 0.1 M) of hydrochloric acid may be used.  In some cases, a 
dilute solution (about 10:1) of water and chlorine bleach or a solution of water and EDTA may also 
work.  Stubborn oil or grease deposits may require cleaning with a solvent such as acetone or 
alcohol.  Protein deposits may be cleaned with a pepsin-based cleaning solution.  Bacterial or mold 
growth may be removed with dilute chlorine bleach.   
 
 
Caution:  Ensure that dangerous chemical reactions will not occur between process 
deposits and cleaning solutions.  Compounds of cyanide and hydrochloric acid will react 
and pose health risks.  Cyanide is often used in electroplating and in gold refining.  If in 
doubt about potential chemical reactions, check with a chemist before cleaning.   
 
4.1.3.  Before returning the sensor to service, allow it to soak in water or buffer at ambient temperature for 
about an hour to restabilize the pH electrode and the salt bridge. 
 
4.1.4.  After cleaning the sensor, calibrate sensor per instructions in the Alpha pH2000D controller manual. 
 
4.1.5.  Reinstall sensor in process.