Calcium Electrode 
 
17
TROUBLESHOOTING HINTS 
 
Symptom             
Possible Causes           
 
Next Step 
Out of Range         
defective meter            
 
check meter with shorting strap 
Reading                                      
 
 
 
(see meter instruction manual) 
 
                       
defective electrode     
 
check electrode operation 
 
                       
electrodes not               
 
unplug electrodes and reset 
                       
plugged in properly 
 
                       
reference electrode       
 
be sure reference electrode is filled 
                       
not filled               
 
 
 
 
                       
air bubbles on             
 
remove bubbles by 
                       
membrane                   
 
re-dipping electrode 
 
                       
electrodes not             
 
put electrodes in solution 
                       
in solution              
 
 
 
 
Noisy or Unstable  
defective meter            
 
check meter with 
Readings (readings                             
 
 
shorting strap 
continuously or 
randomly changing     air bubble on              
 
remove bubble by 
                       
membrane        
 
 
re-dipping electrode 
 
                       
meter or stirrer           
 
ground meter or stirrer 
                       
not grounded               
 
 
 
                       
outer filling solution    
 
fill electrode to level just 
                       
too low                  
 
 
below fill hole 
 
                       
defective electrode       
 
replace electrode 
 
                       
electrode exposed to     
 
soak electrode in calcium standard 
                       
interferences              
 
 
 
"Incorrect           
incorrect scaling          
 
plot millivolts on the linear 
Answer"              
of semi-log paper         
 
axis.  On the log axis, be sure 
(but calibration                             
 
 
 
concentration numbers within 
curve is good)                                                         
 
each decade are  increasing  
                                               
 
 
 
increasing concentration 
 
                       
incorrect sign             
 
be sure to note sign of  
                                            
 
 
 
 
millivolt number correctly 
 
                       
incorrect standards       
 
prepare fresh standards 
 
                       
wrong units used          
 
apply correct conversion factor: 
 
         1.0x10
-3
M = 40 ppm Ca
+2
 = 
                                               
 
 
 
100 ppm as CaCO
3
 
                       
sample carryover          
 
rinse electrodes thoroughly