eutech-instruments bromide epoxy Manuel D’Utilisation

Page de 19
Instruction Manual 
 
Bromide Electrode 
 
15
TROUBLESHOOTING GUIDE 
 
The goal of troubleshooting is the isolation of a problem through checking each of the system 
components in turn: the meter, the glass-ware, the electrodes, the standards and reagents, the 
sample, and the technique. 
 
Meter 
The meter may be checked by following the check-out procedure in instrument instruction manual. 
 
Glass-ware 
Clean glass-ware is essential for good measurement. Be sure to wash the glass-ware well with a 
mild detergent and rinse very well with distilled or deionized water. Clean glass-ware will drain 
without leaving water droplets behind. 
Electrodes 
The electrodes may be checked by using the procedure found in the sections entitled 
Electrode 
Slope Check
 
1. 
Be sure to use distilled or deionized water when following the procedures given in 
Electrode Slope Check
 
2. 
If the electrode fails to respond as expected, see the  sections 
Measuring Hints
 and 
Electrode Response
. Repeat the slope check. 
 
3. 
If the electrodes still fail to respond as expected, substitute another bromide ion electrode 
that is known to be in good working order for the questionable electrode. If the problem 
persists and you are using an electrode pair, try the same routine with a working reference 
electrode. 
4. 
If the problem persists, the standards and/or reagents may be of poor quality, interferences 
in the sample may be present or the technique may be faulty. (See sections 
Standards and 
Reagents
Sample,
 and
 Technique
 below.) 
 
5. 
If another electrode is not available for test purposes, or if the electrode in use is suspect, 
review the instruction manual and be sure to: 
- Clean and rinse the electrodes thoroughly. 
- Prepare the electrodes properly. 
- Use the proper filling solution. 
- Adjust the pH and the ionic strength of the solution by the use of the proper ISA. 
- Measure correctly and accurately. 
- Review 
TROUBLESHOOTING HINTS
 
Standards and Reagents 
Whenever problems arise with the measuring procedure that has been used successfully in the past, 
be sure to check the standard and reagent solutions.  If in doubt about the credibility of any of the 
solutions, prepare them again. Errors may result from contamination of the ISA, incorrect dilution 
of standards, poor quality distilled or deionized water, or a simple mathematical miscalculation. 
 
Sample 
Look for possible interferences, complexing agents, or substances which could affect the response 
or physically damage the sensing electrode (or the reference electrode) if the electrodes work 
perfectly in the standard, but not in the sample.