eutech-instruments cyberscan pc 510 phconductivity meter Manuel D’Utilisation

Page de 56
Instruction Manual                                                                                                                                                                  PC 510 
 
50 
18  ADDENDUM 5: METER FACTORY DEFAULT SETTINGS 
Resetting the meter to factory default settings clears all calibration data and most other setup 
functions you might have changed. The following settings will remain as you have set them: 
•  Temperature unit of measure (°C or °F) 
•  The temperature offset calibration value. 
NOTE: Conductivity and pH data are cleared separately from each other. 
•  To clear pH data, see Sub group P4.0. 
•  To clear conductivity/TDS data, see Sub group P9.0. 
 
Type Parameter 
Default 
Remarks 
pH parameters 
P1.1 
P1.2 
P1.3 
P1.4 
P1.5 
View pH calibration data 
---- 
---- 
---- 
---- 
---- 
No calibration data for 1
st
 buffer, pH 1.68 
No calibration data for 2
nd
 buffer, pH 4.01 
No calibration data for 3
rd
 buffer, pH 7.00 or 6.86 
No calibration data for 4
th
 buffer, pH 10.01 or 9.18 
No calibration data for 5
th
 buffer, pH 12.45 
P2.1 
P2.2 
View electrode offset 
View electrode slope 
0.00 mV 
100.0% 
No offset adjustment 
No slope adjustment 
P3.1 
P3.2 
P3.3 
P3.4 
Ready indicator 
# pH calibration points  
Buffer set option 
°C or °F 
Ready on 

USA 
No default 
Ready indicator on; auto endpoint off 
3 pH calibration points available (1 – 5 pt range) 
USA or NIST buffer set 
°C or °F remains as selected 
P4.0 
pH factory default 
No 
Retains your current settings 
Conductivity and TDS parameters 
P5.1 
P5.2 
P5.3 
P5.4 
P5.5 
Viewing conductivity or TDS 
calibration data 
---- 
---- 
---- 
---- 
---- 
No calibration data for range 1 
No calibration data for range 2 
No calibration data for range 3 
No calibration data for range 4 
No calibration data for range 5 
P6.1 
P6.2 
P6.3 
P6.4 
P6.5 
Viewing conductivity or TDS 
calibration data 
1.0 
1.0 
1.0 
1.0 
1.0 
No offset for effective cell constant for range 1 
No offset for effective cell constant for range 2 
No offset for effective cell constant for range 3 
No offset for effective cell constant for range 4 
No offset for effective cell constant for range 5 
P7.1 
P7.2 
P7.3 
P7.4 
Ready indicator 
°C or °F 
ATC on or off 
TDS factor 
Ready on 
No default 
ATC on 
0.5 
Ready indicator on; auto endpoint off 
°C or °F remains as selected 
---- 
Adjustable from 0.4 to 1.0 
P8.1 
P8.2 
Temperature coefficient 
Normalisation temperature 
2.1% per 
°C
25 
°C 
Adjustable from 0 to 10% 
Adjustable from 15 to 30 
°C 
P9.0 
Conductivity factory default 
No 
Retains your current settings