eutech-instruments ammonia gas Manuel D’Utilisation

Page de 21
Instruction Manual 
 
Ammonia Gas-sensing Electrode
 
 
16
TROUBLESHOOTING HINTS 
 
Symptom                  Possible Causes          
 
Next Step 
Out of Range    
defective meter 
 
 
perform meter checkout procedure 
Reading                                      
 
 
 
(see meter instruction manual) 
 
                          
defective inner body         
 
refer to 
Checking the Electrode 
                                     
 
 
 
Inner Body 
 
                          
electrode not           
 
 
unplug electrode and reset 
                          
plugged in properly  
 
                          
internal filling        
 
 
fill outer body of electrode with proper 
                          
solution not added        
 
amount of internal filling solution 
 
                          
air bubble on membrane          
 
remove  bubble  by  re-dipping 
electrode 
 
                          
electrode not in solution 
       
put electrode in solution 
                         
Noisy or Unstable     insufficient            
 
 
fill outer body of electrode with proper  
Reading (readings      internal filling solution 
 
amount of internal filling solution 
continuously or                          
randomly changing.)   defective meter           
 
perform meter checkout procedure 
                                                 
 
 
 
(see meter instruction manual) 
 
                          
bottom cap loose          
 
ensure that bottom cap is screwed on 
                                                 
 
 
 
tight enough to close gap between 
                                                 
 
 
 
bottom cap and body 
 
                          
defective inner body        
 
refer to 
Checking the Electrode 
                                         
 
 
 
 
Inner Body 
 
                          
air bubbles on membrane         
remove bubbles by redipping electrode 
 
                          
meter or stirrer        
 
 
check meter and stirrer for 
                         
improperly grounded     
 
grounding 
 
 
Drift (reading slowly  internal filling        
 
 
ensure that membrane is installed 
changing in one          solution leakage             
 
properly 
 
                          
incorrect internal        
 
refill outer body of electrode using      
                       
filling solution        
                   
filling solution shipped with electrode 
 
                          
total sample level        
 
dilute sample of dissolved species 
                          
 
 
 
 
 
above 1M 
 
                          
electrode in sample      
 
reduce surface area to volume ratio, 
                          
too long; NH
3
 loss       
 
slow down rate of stirring, avoid