Setting the
Standard
in Water
Testing
Built in
temperature
sensor for ATC
Rugged epoxy
housing with
1m cable
68X247782 12/01 Rev 0
OPERATING INSTRUCTIONS
Thank you for selecting an Eutech Instruments
conductivity/TDS cell. Cell constant (K) = 1.0
Dimensions (mm): 132 L x 12 dia.
BEFORE YOU BEGIN: 
Soak electrode in alcohol
for a few minutes to remove oils, if any.
CALIBRATION: 
Calibrate according to meter
instructions.  ALWAYS use fresh standard:
see selected Eutech calibration standards
below. Select a calibration value equal to
2/3 of the value in the range you are
measuring.
DURING MEASUREMENT:
• Avoid particulate matter: allow it to settle
before taking measurements. Don't let
sediment build up on the sensors.
• Avoid air bubble entrapment in the sensor
by gently tapping electrode to dislodge.
• For accurate readings, make sure
conductivity cell sensors are completely
immersed in the sample.
• Rinse with deionized water after every
measurement.
CLEANING: 
Soak cell in alcohol for 2 minutes
& then rinse thoroughly in deionized water.
ACCESSORIES:
Conductivity Solutions (480ml per bottle)
EC-CON-1413BT 1413 µS KCI Calibration
Solution
EC-CON-1288BT 12.88 mS KCI Calibration
Solution
EC-442300BT
300 ppm 442 Calibration
Solution
EC-4423000BT
3000 ppm 442 Calibration
Solution
Contact your distributor for more solutions.
WARRANTY:
 This cell is warrantied for a
period of 6 months from the date of purchase.
If  repair, adjustment or replacement is
necessary and has not been the result of
abuse or misuse within the 6 month period,
please return the cell – freight pre-paid –
and correction will be made without charge.
Out of warranty products will be repaired on
a charge basis.
RETURN OF ITEMS: 
Authorization must be
obtained from your distributor before
returning items for any reason.
Note: We reserve the right to make changes, improvements
and modifications to products shown without prior notice.