eutech-instruments lead epoxy Manuel D’Utilisation

Page de 23
Instruction Manual 
 
Lead Electrode
 
 
19
TROUBLESHOOTING HINTS 
 
Symptom                 Possible Causes         
 
Next Step 
Out of Range            defective meter          
 
check meter with shorting strap  
Reading                                      
 
 
 
(see meter instruction manual) 
 
                         
defective electrode      
 
check electrode operation 
                                             
                         
electrodes not         
 
 
unplug electrodes and reset 
                        
plugged in properly 
 
                         
reference electrode       
 
be sure reference electrode 
                         
not filled         
 
 
is filled 
 
                         
air bubble on membrane          
remove bubble by re-dipping electrode 
 
                         
electrodes not        
 
 
put electrodes in solution 
                         
in solution            
 
 
Noisy or               
defective meter          
 
check meter with shorting strap 
Unstable Readings                            
(readings  
 
air bubble on membrane 
 
 remove bubble by re-dipping electrode 
continuously                  
or rapidly changing)  electrode exposed        
 
soak electrode in lead standard 
                         
to interferences       
 
                         
defective electrode      
 
replace electrode 
 
                         
ISA not used           
 
 
use recommended ISA 
 
                         
meter or stirrer       
 
 
ground meter or stirrer 
                         
not grounded          
 
 
Drift (reading         
samples and standards  
 
allow solutions to come to room 
slowly                 
at different temperatures        
temperature before measurement 
changing in one        
direction)          
complexing agents       
 
check section entitled 
                         
in sample                
 
 
Precipitation and Complexation
 
 
                         
incorrect reference            
 
use recommended 
                         
filling solution     
 
 
filling solution 
                       
                        
membrane dirty           
 
polish membrane; use 
                         
or oxidized            
 
 
methanol-formal dehyde solution