eutech-instruments silversulfide epoxy Manuel D’Utilisation

Page de 26
Instruction Manual 
 
Silver/Silver Sulfide Electrode 
 
19
 
 
50 
  
32.1 
 64.1 
 
If changes in temperature occur, the electrodes should be re-calibrated. The temperature range for  
Eutech Silver/Sulfide Ion Combination Electrode is 0o-80oC, provided that temperature 
equilibrium has occurred. If the temperature varies substantially from room temperature, 
equilibrium times up to one hour are recommended. 
 
Electrode Response 
 
Plotting the electrode mV potential against the silver concentration on semi-logarithmic paper 
results in a straight line with a slope of about 57 mV per decade. (Refer to Figure 2.) The sulfide ion 
also gives a straight line when the electrode mV potential is plotted against the sulfide 
concentration, but the slope is about 26 mV per decade.  (See Figure 1.) 
 
The time needed to reach 99% of the stable electrode potential reading, the electrode response time, 
varies from several seconds in highly concentrated solutions to several minutes near the detection 
limit. (Refer to Figure 5.) 
 
 
A drifting potential reading or a decrease in electrode slope may mean that the electrode membrane 
needs polishing. 
 
To polish the membrane: 
 
1. 
If using polishing paper, cut off a 1-2" piece and place it face up on the lab bench. 
 
2. 
Put a few drops of distilled or deionized water in the center of the paper. 
 
3. 
Holding the paper (cotton) steady with one hand, bring the membrane of the electrode 
down perpendicular to the paper and, with a slight swirling motion, gently polish the tip of 
the electrode against the surface of the polishing paper (cotton) for a few seconds. 
 
4. 
Rinse the electrode surface with distilled or deionized water and soak the electrode tip in 
standard solution for about five minutes before use.