eutech-instruments waterproof c401 colorimeter Manuel D’Utilisation

Page de 44
Measurement: pH
 
 
C401/301/201/102/101 
 
 
 
20
Invert several 
times to mix. 
Add 20 drops of pH Indicator
(Phenol Red) reagent 
     pH
Indicator
  Phenol
    Red
 
9. 
Gently push the vial back fully into the sample well of the instrument, 
whilst again aligning the ▼ mark on the vial with the ▲ mark on the 
meter. 
10.  Wait for 1 minute and then press the READ/ENTER key within 3 
minutes from when the time the reagent was added to the vial. 
11.  The meter will flash [PH] whilst the measurement is taken.  It will then 
display the pH of the solution in pH units. 
       pH 
   Sample
READ
ENTER
pH