Restoring Factory Calibration
 
 
C401/301/201/102/101 
 
 
 
33
3.5 
Restoring Factory Calibration 
The C401/301/201/102/101 Colorimeter allows you to reset the meter 
back to the factory default calibration values. This feature is extremely 
useful when there are errors in calibration or when you have new 
calibration standards.  
1. With the meter switched off, press MODE and ON/OFF keys 
simultaneously for about 2 seconds. 
2.  The meter will momentarily flash the model number and [URST] (User-
Reset) and then display [No]. 
3.  Use MODE key to select between [YES] or [No].  
ƒ
  [YES] = To restore meter back to factory calibration values 
ƒ
  [No] = To retain last calibrated values 
4.  After selecting [YES] or [No], press READ/ENTER key.  
5. The meter will now flash the model number, the meter’s revision 
number and enter the measurement selection mode. 
Figure 9 shows the sequence for restoring factory calibration values.