eutech-instruments waterproof c401 colorimeter Manuel D’Utilisation

Page de 44
Measurement
 
 
C401/301/201/102/101 
 
 
 
6
2 MEASUREMENT 
2.1 
General Information 
The waterproof C401/301/202/102/101Colorimeter allows you to measure 
up to four parameters - namely the pH, Cyanuric acid, Free Chlorine and 
Total Chlorine of a sample and reports the results in appropriate units 
(ppm or pH). Each specific measurement has its own range as listed in the 
specifications. 
The meter will go through a power sequence as shown below.  
Meter’s Model Number
Meter’s Revision Number
Measurement Mode Selection 
 
Figure 5: Power Up Sequence 
An accurate colorimeter measurement depends on good measurement 
techniques. Factors such as clean sample vials, positioning of vial in the 
sample well, covering the vial with the light shield cap, meter calibration, 
handling of meter, and others, have to be taken into consideration.  
NOTE: Do not pour liquid directly into the sample well of the instrument. 
Always use a vial. The instrument will only accurately measure the sample 
when vials sealed with the black caps are used. The black cap serves as 
both seal and a light shield. 
NOTE: Do not attempt to clean the sample well. The optics may be 
damaged. 
NOTE: This instrument is designed to measure solutions contained in the 
custom made glass vials and therefore the use of other type of sample vials 
or sample cells may result in inaccurate measurements.