eutech-instruments field phorpconductivity monitors Manuel D’Utilisation

Page de 19
 
 
10
8.2  pH/ORP electrode cleaning 
•  Salt deposits: Dissolve the deposits by immersing the electrode in tap water for 
ten to fifteen minutes. Then thoroughly rinse with distilled water.  
•  Oil/Grease film: Wash electrode pH/ORP bulb gently in some detergent and 
water. Rinse electrode tip with distilled water or use a general-purpose electrode 
cleaner (see ordering information).  
•  Clogged reference junction: Heat a diluted KCl solution to 60 to 80  C. Place the 
sensing part of the electrode into the heated solution for about 10 minutes. Allow 
the electrode to cool in some unheated KCl solution.  
8.3  Conductivity electrode  
For general cleaning, the use of deionized water will be sufficient.  Rinse the 
electrode thoroughly in the deionized water.  After cleaning, wipe the electrode dry.  
Never touch the metal sensors.  
For intensive cleaning, when the electrode is contaminated, wash it carefully with a 
mild detergent, after which wash thoroughly with deionized water.  Always ensure 
that there are no blockages of the electrode cavity.  If there is, use a gentle jet of 
water to dislodge any particles that may have been stuck there.   
 
9 TROUBLE-SHOOTING 
GUIDE 
Problem Cause 
Solution 
No display 
a). Power adapter is not  
connected properly or is 
faulty.  
 
 
a). Check that power adapter and 
its wire are in place and 
making good contact.  
b). Check power supply.   
Unstable 
readings 
a). Air bubbles in probe. 
 
b). Dirty probe.  
 
c). Probe not deep enough 
in sample.  
d). External noise pickup or 
induction caused by 
nearby electric motor.  
a). Tap probe to remove bubbles.  
 
b). Clean the probe and re-
calibrate.  
c). Make sure sample entirely 
covers the probe sensors.  
d). Move or switch off interfering 
motor.  
 
No response 
a). Broken probe. 
a). Replace probe 
Slow response 
a). Dirty / Oily probe.  
a). Clean probe. See “Electrode 
Care & Maintenance” section.