INTERCONNECTIONS KIT, P/N 401699, 
INCLUDES 
QTY PART 
NO 
FIGURE 
CALLOUT DESCRIPTION 
401697 
N/A 
Dual Volume Unit (DVU) 
101450-Z 
Tubing, PFA, 1/4 in. x 160 cm 
2 103279 
Adaptor, 1/4 in. tube x O-seal 
1/8 in. NPT M 
4 101477 
Adaptor, 1/4 in. tube x 1/4 in. 
NPT M 
2 101985 
4  Union, 
4P 
101487 
Tee, 1/4 in. tube 
2 101482 
Adaptor, 1/4 in. tube x 1/4 in. 
NPT F 
 
TO INSTALL THE DUAL VOLUME UNIT (DVU) 
  Identify a location for the DVU™ that minimizes 
the length of tubing between the DVU connections 
and the PPC BG0002™ TEST(+) and TEST(-) ports.  
The PPC BG0002 and the DVU are designed to 
stack one on top of the other.  The DVU may be 
rack mounted using the rack mount kit (P/N 400440).  
Recommended configurations are for the PPC and 
DVU to be stacked directly one on top of the other 
or with their rear panels face to face. 
TO CONNECT THE DVU AND PPC BG0002 
 
 
You will be making and connecting two identical 
assemblies, one for the TEST(+) port and one for 
the TEST(-) port. 
 
  Install a union (4) into a 1/4 in. tube x 1/4 in. NPT 
M adaptor (3) (do this twice).  Screw the 1/4 in. 
NPT M end of the adaptors into the two ports of 
the DVU using Teflon
®
 tape as a sealer. 
 
 
Figure 1  Interconnections Kit 
 
  Install a 1/4 in. tube tee (5) onto each of the 
union/adaptor assemblies that were just installed 
into the DVU ports in Step  .  Orient the tee 
considering the connection to the PPC BG0002 
(schematic above is for PPC stacked on top of  
the DVU). 
  Install the appropriate adaptors onto the PPC BG0002 
TEST(+) and TEST(-) ports:  
For PPC3™  Install the 1/4 in. tube x O-seal  
1/8 in. NPT M (2). 
For PPC2+™  Install the 1/4 in. tube x 1/4 in.  
NPT M adaptor (3) using Teflon 
tape as a sealer. 
  Cut lengths of PFA tubing as needed to connect 
the 1/4 in. tube tee to the adaptors on the PPC 
TEST ports.  Take care that the cut is at a right 
angle to the tube and only cut the length needed. 
  Optional for a connection “For DUT” of 1/4 in. NPT M 
rather than 1/4 in. tube: 
•  Cut a small length of PFA tubing to use as  
a union between the 1/4 in. tube tee (5) and  
the 1/4 in. tube x 1/4 in. NPT F adaptor (6) 
(see schematic above).