dhi pk-7000-mpc1-d interconnections kit installation instructions Manuel D’Utilisation

Page de 2
 
Calibration Solutions 
for Pressure 
and Flow™ 
PK-7000-MPC1-D 
Interconnections Kit 
Installation Instructions 
INTERCONNECTIONS KIT, P/N 401645, INCLUDES 
• 6 100319 Adaptor, 1/8 in. tube x 1/8 in. NPT M 
• 2 102184 
Reducer, 4P x 2P 
• 2 100317  Adaptor, 1/8 in. tube x 1/4 in. NPT F 
• 1 101988-Z  Tubing, SS, .125 in. OD x 300 cm 
• 2 401459  Quick Connect and Stand Assy 
• 2 102116 
Quick Connect, 1/8 in. NPT F 
PARTS ON TEST PORT OR IN ACCESSORIES OF PG7000 
• 1 101200 Collar 
• 1 400518 Adaptor , 1/8 in. tube x DH 200 M  
• 1 100272 Gland 
 
 
 
 
 
TO PNEUMATICALLY CONNECT THE PG7000
S
 OR RPM
S
 TO THE MPC1-D 
• 
MPC1-D is intended for controlling pressures in a high line differential pressure system where there is a “high” side and 
a “low”.  The instructions below should be repeated, once for the high side and once for the low side.  See the MPC1-D 
Operation and Maintenance manual for a schematic of a typical system layout. 
 
• 
If you are using RPMs (not PG7000s) go directly to Step 
p
b. 
 
Remove gland, collar and plastic plug from the PG7000 TEST port. 
Install gland and collar onto 1/8 in. tube x DH200 male adaptor supplied in PG7000 accessories.  
Note that the collar threads left handed.  Assure it is properly attached.  The orange plastic plug is 
not used. 
pa  PG7000 - Screw the adaptor (as assembled) into the PG7000 TEST port.  Torque to 15 Nm (Newton 
meter). 
pb  RPM3 - Screw 1/8 in. tube x 1/8 in. NPT male adaptor into the RPM3 TEST port using Teflon tape as 
a seal. 
TO MPC1-D
TEST HI or TEST LO
TO MPC1-D
REF HI or REF LO
101988-Z
102184
100319
100319
100317
102116
401459
101988-Z
400518
INCLUDED WITH P 7000
G
100272
100319
TO RPM3
101200
TO P 7000
G