WAVE PLUS - USER MANUAL 
C
OD
.
 
WPMEMG
 
REV 
01 
02/11/2011
 
D
ATA PRIMA EMISSIONE
 
 
Pag. 18 di 37 
 
event  of  the  beginning  and  the  end  of  the  data  acquisition  or,  in  general,  a  temporal  reference  event 
(e.g. the timing of the TMS stimulation) can be directly saved into the same data streaming. 
 
The scheme above shows the required connections to realise this type of synchronism in case: 
1.  The third-party device is working as master: red connection; 
2.  The Wave Plus system is working as master: blue connection
1
 
 
 
NOTE THAT if you want synchronize the Wave Plus system with third-party devices collecting the 
EMG signals from the Wave Plus analogue port, you should also consider the 13 ms delay related to 
the Wave Plus analogue output. In this case you should correct the recorded data in order to remove the 
13 ms delay. NOTE THAT the 14 ms delay is constant and it can be easily removed by shifting the 
temporal sequence in the following way: 
 
 
 
 
 
 
 
 
 
 
NOTE THAT the delay does not produce a loss of acquired data at all. 
Please, for further details contact Cometa. 
 
4.2.2  Supported external devices 
 
Wave Plus system can be connected to any PC with USB interface; these are the features required by 
systems: 
  Type 2 USB interface  
  Windows 2000 or XP  
  VGA graphic card with a resolution of at least 1024 x 768 
  Minimum ram memory, 512 MB 
  Minimum processor frequency, 800 MHz 
  Minimum hard disk, 20 GB 
 
 
                                                      
1
 In this case, it’s necessary to use also the dedicated software in order to produce the trigger signal. 
Unrecorded sequence 
Recorded sequence 
Unrecorded sequence 
Trigger Off 
Trigger On 
Trigger Off 
Unrecorded sequence 
Recorded sequence 
Unrecorded sequence 
14ms