Cometa Sas di A. Dellacorna & C. WPTX Manuel D’Utilisation
WAVE PLUS - USER MANUAL
C
OD
.
WPMEMG
REV
01
02/11/2011
D
ATA PRIMA EMISSIONE
Pag. 21 di 37
If the batteries are deeply discharged, it may be necessary to perform two or more charge
cycles, to restore its original capacity.
cycles, to restore its original capacity.
The electrode, when not in use, has a very low power consumption and maintains the charge for a long
time, but to maximize the life of the battery, in case of non-use, we suggest to perform at least one
charge cycle every three months.
time, but to maximize the life of the battery, in case of non-use, we suggest to perform at least one
charge cycle every three months.
4.5.2 LED indicator of the electrodes
When the probes are charging, the green LED light provides an indication about the level of charge.
Particularly, three states are available:
Particularly, three states are available:
1) LED orange flashing at one flash/sec.: probe on pre-charge (when battery is entirely un-
charged);
2) LED orange flashing at five flash/sec.: probe on normal charging phase;
3) LED orange off: full charge
The transmitter units can be charged by placing on the charger the only lower part of the electrode. The
feedback on the state of charge takes place through a second green LED placed within that module.
Charging is complete when the LED is off.
IMPORTANTNOTE. The unit placed in the docking box, during the charging process, do not trans-
mit. This is valid only during the charging process (charging module active -Orange LED). At the end
of the charging cycle, the transmitting units resume its normal functionality.
feedback on the state of charge takes place through a second green LED placed within that module.
Charging is complete when the LED is off.
IMPORTANTNOTE. The unit placed in the docking box, during the charging process, do not trans-
mit. This is valid only during the charging process (charging module active -Orange LED). At the end
of the charging cycle, the transmitting units resume its normal functionality.
4.6 Warnings
Extended non-use of the device
If the device has not been used for a long time, we suggest to:
Recharge the electrodes battery every six months, or,
Unlink/Disconnect the EMG acquisition module from the battery module
and keep the elements in a dry and clean environment.
How to use the device in an environment in which a wireless LAN is working
If the Wave Plus receiving unit is close to an active WLAN unit, some interference
could occur causing a loss of EMG data. Moving away the two units, the system
would work properly, at least one or two meters. Data loss is shown by the
corresponding “CH#” led that temporary switches off. Data loss is always
recognizable. In case of continuous loss of data and/or problems of acquisition, it
could occur causing a loss of EMG data. Moving away the two units, the system
would work properly, at least one or two meters. Data loss is shown by the
corresponding “CH#” led that temporary switches off. Data loss is always
recognizable. In case of continuous loss of data and/or problems of acquisition, it