Appendix B – Calibration, Traceability, and Recalibration  
Ophir Thermopile Heads 
Accuracy of Calibration 
Since the instruments are calibrated against NIST standards, the accuracy is 
generally 1% at the power level at which the calibration has been performed. 
This accuracy has been verified by checking the scatter of the results when 
several instruments are calibrated against the same standard. The maximum 
measurement error is less than the sum of the specified accuracy and linearity. 
Since the linearity is also 1%, the maximum error in measurement is generally 
less than 2%. 
Recalibration from a Known Source of Laser Power
 
3A Series 
1 Watt 
7 Volts 
10A Series 
10 Watts 
17 Volts 
30A Series 
30 Watts 
35 Volts 
150A Series 
100 Watts 
65 Volts 
L250A Series 
200 Watts 
95 Volts 
300W Series 
125 Watts 
65 Volts 
1500W Series 
and above 
200 Watts 
95 Volts 
1.  Click the right mouse button and select Calibrate from the pop-up menu 
OR 
Open the Functions menu and select Calibrate. The Adjust Calibration 
Factors
 dialog box appears. 
 
Figure B-3 Adjust Calibration Factors Dialog Box 
The Original area displays the original MeasurementOverall Sensitivity 
and Laser Factor fields. The Current wavelength is displayed beneath the 
Original area, refer to Note on page 97. 
StarLab User Guide 
96