Appendix B – Calibration, Traceability, and Recalibration
Ophir Photodiode Heads
Table B-2 Maximum Error as a Function of Wavelength and Filter
Wavelength
Error, Filter Out
PD300
PD300 -
3W
3W
PD300 -
UV
UV
PD300 - IR 3A - IS
220-250nm
±6%
250-360nm
±3%
360-400nm ±10% ±10% ±3%
400-950nm ±3% ±3% ±3%
±5%
950-1100nm
±5% ±5% ±5% ±4% ±10%
1100-1700nm
±4%
1700-1800nm
±7%
Note: Add ±2% to error for filter in (±4% for PD300-UV from 220 to
300nm).
User Recalibration
Adjusting Calibration Factors of Photodiode Heads
To adjust the power calibration factor:
1. Click the right mouse button and select Calibrate from the pop-up menu.
OR
Open the Functions menu and select Calibrate. The Adjust Calibration
Factor dialog box appears.
Factor dialog box appears.
Figure B-5 Adjust Calibration Factor Dialog Box
The Original area displays the original Measurement and Calibration
Factor. The Current wavelength is displayed beneath the Original area.
Factor. The Current wavelength is displayed beneath the Original area.
Note: A different wavelength can be selected in the Measurement Parameters
Area.
StarLab User Guide
99