Appendix B – Calibration, Traceability, and Recalibration  
Ophir Photodiode Heads 
Table B-2 Maximum Error as a Function of Wavelength and Filter 
Wavelength 
Error, Filter Out 
 PD300 
PD300 - 
3W 
PD300 - 
UV 
PD300 - IR  3A - IS 
220-250nm 
  ±6% 
  
250-360nm 
  ±3% 
  
360-400nm ±10% ±10% ±3%   
 
400-950nm ±3% ±3% ±3%  
±5% 
950-1100nm 
±5% ±5% ±5% ±4% ±10% 
1100-1700nm 
   ±4% 
 
1700-1800nm 
   ±7% 
 
Note: Add ±2% to error for filter in (±4% for PD300-UV from 220 to 
300nm). 
User Recalibration 
Adjusting Calibration Factors of Photodiode Heads 
To adjust the power calibration factor: 
1.  Click the right mouse button and select Calibrate from the pop-up menu.  
OR 
Open the Functions menu and select Calibrate. The Adjust Calibration 
Factor
 dialog box appears. 
 
Figure B-5 Adjust Calibration Factor Dialog Box 
The Original area displays the original Measurement and Calibration 
Factor
. The Current wavelength is displayed beneath the Original area. 
Note: A different wavelength can be selected in the Measurement Parameters 
Area. 
StarLab User Guide 
99